Oxygen Analyzer
T-O2ZBA-25

Oxygen Analyzer

Oxygen analyzer adopts imported new long-life imported zirconia detector with high sensitivity and fast response. Nitrogen-oxygen analyzer has a high-precision temperature automatic compensation system to eliminate the influence of ambient temperature.

Related Products

Dissolved... Dissolved...

The imported new-type longevous zirconia detector is with high accuracy and a rapid response speed for features, etc;

N2 or O2 display mode is available;

The auto-compensation system for temperature with high accuracy could eliminate the influence from the environment temperature;

The upper and lower alarm limits can be set freely among the whole range;

The accurate measurement of oxygen concentration among the range of ppm~% can be met through one-point calibration in the air;

With power supplied at a wide range, the influence from mains fluctuation on the analyzer can be eliminated;

With an external printer, the test parameters can be printed manually or automatically (optional);

The size is small and easy to maintain.

It is widely used in N2 reflow furnace, N2 direct flow furnace, air separation device, gas purity management, protective inert gas, ferrite sintering furnace and other environments to measure oxygen concentration

Parameter

Description

Test principle

zirconia

Display mode

5 bits LED display

Test range

10/100/1000ppm,21.000% O2/N2

Accuracy

≤±1.5% FS

Repeatability

≤±1%

Response time

T90≤30S

Analog output

4-20mA.DC(non-isolated output, loaded

 resistance lower than 500 Ohm)

 1 way programmable alarm output (Dr y

 contact type, maximum load 220VAC/2A)

Environment humidity

<90%RH

Power supply

AC85~264V 50/60Hz

Sample gas temperature

-10 ~ +50℃

Sampling mode

sample gas pass-in

Sample gas flow

300 ~ 500ml/min

Background gas

nitrogen and oxygen mixed gas

Appearance Size

80mm×160mm×142mm(H×W×D)

Install Hole Cut

75mm×151mm(H×W)

Serving life

20000 hours(under normal use)

Gas cycle port

Φ6 plastic pipe inlet and outlet

Install mode

frame mount or retaining clap carried by the

 oxygen sensor