Scanner - Envada

Description

The EN6000B6 series instrument is a multichannel measurement instrument, which is offered with two alternative display types: Digital or OLED. It is suited for measurement, display and alarm multi-channel signals at the place of work. The instrument can receive varieties of signal including standard current, standard voltage, standard thermal resistor, standard thermocouple, etc; No jumper is needed; setup of the internal parameters is all that is needed using pre-selected options between a varieties of common signals.

A RS485 communication interface is available (as an option) to form a computer monitoring system based on multiple instruments and a PC that has the bus characteristics defined by the communication protocol (supported by many host computer configuration software) or MODBUS protocol allowing the functional operation of a instrument to be carried out on a host computer seamlessly. The user can choose a communication rate among 2400, 4800, 9600, 19200bps. The standard communication software operated in a PC is available.

  • Intelligent multi channel data acquisition/ display
  • Universal signal input function
  • 8/ 16 Channels - Universal Input (TC/ RTD/ mA/ mV/ V)
  • Every channel is independently programmable for input, range & alarms
  • Input channels are isolated for all inputs except RTD
  • Communication (RS 485/RS232) with software up to 4 Alarms, 3 Inputs, I/O
  • Common/ Independent Low & High Alarm relays for each channel
  • 24VDC Power supply for 2 wire transmitters – on 3 channels
  • High accuracy (14 bit resolution), µP based
  • Self Diagnosis function: Input fail safe operation
  • Extended alarm function.
  • Advanced parameter protection function can be set to avoid any illegal operation
  • Flow integration with temp & Pressure compensation for steam & supersaturated steam
  • EMI/ ESD protection; CE approved

PARAMETERS

SPECIFICATIONS

INPUT

Thermocouple

S, B, T, R, K, E, J, N

RTD

Pt100, Cu50, Cu100

Voltage

(-)20 - 20mV DC, (-)100 -100mV DC, 0 -100mV DC, 0 - 500mV, 0-1VDC, 0-5V DC, 1-5V DC, 0-10V DC

Current Signals

0-10mA, 4-20mA, 0-20mA

Square root extraction

0-5V DC, 1-5V DC, 0-10mA, 4-20mA

Resistance Signals

0-100 Ω, 30-350 Ω

MEASURING ACCURACY

± 0.2% FS ±1 Digit, typical

RESOLUTION

T/C: 1°C (0.1°C upto 999.9°C) / RTD: 0.1°C

DISPLAY

Process Value

Type

4 digit, 7 segment LED, Red

Size

14 mm (0.56”) in 96 x 96 mm; 9 mm (0.35”) in 48 x 96 or 96 x 48

Channel No.

Type

4 digit, 7 segment LED, Green

Size

9 mm (0.35”)

CHANNEL DISPLAY TIME

Programmable from 1 to 100 seconds

OUTPUT (Optional)

Linear Retransmission output

4-20mA (14bit resolution), isolated, 0-20mA (option)

Accuracy

± 0.3% FS, typical

Alarm output

None/ Common One/ Two (High / Low) Alarm Relay/s (built-in) / Independent High & Low Alarm Relays per channel (in external wall mount enclosure);Option: SSR drive Voltage in place of Relay

Alarm relay rating

Common alarm: 1 Amp/ 250 VAC, resistive load; 3/10 A (Optional) Independent alarm: 3 Amp/ 250 VAC, resistive load

TRANSMITTER POWER SUPPLY

24V DC/ 30 mA, isolated for 2 wire type; upto 3 nos. (Optional)

COMMUNICATION (Optional)

RS485 - ASCII Protocol/ MODBUS / Serial Printer Port)

SIZE

Bezel (W x H), mm

96 x 96 or 48 x 96 or 96 x 48

Panel cut out (W x H), mm

90.5 x90.5 44.5 x 90.5 90.5 x 44.5

 

  • Ovens
  • Furnaces
  • Ceramic kilns
  • Boilers
  • Heat Exchangers

 